-150°C Cryogenic Chest Freezer
Solicitation number B0167
Publication date
Closing date and time 2021/08/26 14:00 EDT
Description
Background
The National Centre for Foreign Animal Disease (NCFAD) in Winnipeg has an urgent need to
replace a failed -150°C freezer. The freezer is used to preserve risk group 4 agents for research
and diagnostic preparedness for high consequence foreign animal diseases of high economic
impact and high danger to Canadians. The stocks include Ebolavirus, Nipah virus, Hendra virus
and other RG4 group viruses. The loss of these stocks would cause great harm and prevent a
timely response to a animal diagnostic response in Canada.
Purpose
This Statement of Work establishes the minimum technical requirements for this procurement for
the acquisition of one (1) -150°C chest freezer.
In order to protect and ensure proper storing conditions of main stocks of viruses used at the
NCFAD, there is an urgent need to replace failed existing freezer.
General requirements
The freezer must be new. A demonstration or refurbished model will not be considered. The unit
must be available for immediate delivery to CFIA warehouse in Winnipeg
The unit must meet or exceed specifications below
The quote must include shipping costs and a minimum five year warranty on parts and labor.
The equipment must be delivered before June 30, 2021.
Technical specifications
The unit has to fit in footprint 68.5” wide and 36.5” deep. Unit cannot be higher 40”
The capacity of freezer must be at least 230 L.
Must have an ability to connect to LN2 backup.
Must be able to maintain temperature at -150°C. The adjustments must be in 1°C
increments.
Must have remote alarm contacts (N.O. & N.C. type, max contact rating -DC 30V, 2
Amp) which allows for hook-up to auxiliary central alarm monitoring systems.
Must have an adjustable temperature alarm points, with visual and audible alarms,
backed up by rechargeable batteries
Must have an access port for adding secondary temperature probes.
Must have casters to facilitate moving and at least two (2) adjustable levelling feet.
Must have lockable latch.
Must have CFC free refrigerants.
Must have a display informing about current temperature, set points, warning messages,
alarms.
Must operate on 208-230V 60Hz single phase 15A current and use 6-15 (NEMA type)
plug.
The operations manual must be provided in paper or electronic format in English.
The device must be CSA approved or equivalent for use in Canada.
Contract duration
Refer to the description above for full details.
Trade agreements
-
Canadian Free Trade Agreement (CFTA)
Contact information
Contracting organization
- Organization
-
Canadian Food Inspection Agency
- Address
-
1400 Merivale RoadOttawa, Ontario, K1A 0Y9Canada
- Contracting authority
- Githua, Virginia
- Phone
- 343-572-4182
- Email
- Virginia.Githua@inspection.gc.ca
- Address
-
59 CamelotOttawa, ON, K2G 5X4CA
Buying organization(s)
- Organization
-
Canadian Food Inspection Agency
- Address
-
1400 Merivale RoadOttawa, Ontario, K1A 0Y9Canada
Bidding details
Full details regarding this tender opportunity are available in the documents below. Click on the document name to download the file. Contact the contracting officer if you have any questions regarding these documents.
Document title | Amendment no. | Language | Unique downloads | Date added |
---|---|---|---|---|
b0167_-_npp_french.pdf |
French
|
4 | ||
b0167_-_npp_english.pdf |
English
|
24 |
Access the Getting started page for details on how to bid, and more.